Description
Dimensions: |
Height: 10.10 in (25.7 cm)
Width: 6.5 in (16.5 cm)
Depth: 4.5 in (11.4 cm)
|
||||
Weight: |
Device: 6.6 lbs (2.98 kg)
Shipping: 12.15 lbs (5.5 kg)
|
||||
Electrical: |
AC: 100 – 240 VAC (1.0 A @ 120 VAC)
DC: 11 – 16 VDC (5.0 A @ 12 VDC)
Internal Battery Average Charge Time: 4 Hours
|
||||
Oxygen Concentration: | 87 to 95% @ 5 l/min setting within 4 to 6 minutes | ||||
Environmental Conditions: |
Temperature Range: 41°F to 104°F (5°C to 40°C)
Storage Temperature: – 4°F to 140°F (- 20°C to 60°C)
Relative Humidity: Up to 95% Non-condensing
Altitude: Sea Level to 14,000 ft. (4,267 m)
|
||||
Outlet Temperature: | Within 2°F (1.1°C) of ambient | ||||
Noise level: | 44 dBA @ setting 2 | ||||
Pulse Settings: |
Setting
1
2
3
4
5
|
Volume
12cc ± 2cc
19cc ± 3cc
26cc ± 4cc
33cc ± 5cc
39cc ± 6cc
|
Peak Flow (l/min)
8.6 ± .3
9.9 ± .2
10.5 ± .3
11.4 ± .2
11.2 ± .2
|
Minute Volume
240 ± 15%
380 ± 15%
520 ± 15%
660 ± 15%
780 ± 15%
|
Battery Duration
4.7 hrs
3.4 hrs
2.7 hrs
2.0 hrs
1.6 hrs
|
External Battery: |
Setting:
1
2
3
4
5
|
Run Time:
4.7 hrs
3.4 hrs
2.7 hrs
2.0 hrs
1.6 hrs
|
Average Charge Time: 3.5 hrs | ||
Trigger Delay: |
Sensitivity: -0.4 cmH20 @ 20 BPM
Delay Time: ≤0.12 seconds @ 20 BPM
|
||||
Cannula: | Maximum 7 ft long standard adult single lumen nasal cannula | ||||
Equipment Classification: |
Class II -with respect to protection from electrical shock
Type BF -degree of protection against electrical shock
IP21 -degree of protection against ingress of solids/liquids
|
||||
Mode of Operation: | Continuous |
*Specifications are subject to change without prior notice.